Influence of Implant Scanbody Wear on the Accuracy of Digital Impression for Complete-Arch: A Randomized In Vitro Trial
Articolo
Data di Pubblicazione:
2022
Abstract:
: The aim of this study was to evaluate the influence of implant scanbody (ISB) wear on the accuracy of digital impression for complete-arch. A polymethylmethacrylate (PMMA) edentulous mandibular model with four internal hexagonal interlocking conical connections was scanned with an extraoral optical scanner to achieve a reference file. Four cylindrical polyetheretherketone (PEEK) ISBs were scanned 30 times with IOS, and the test files were aligned to the reference file with a best-fit algorithm. For each analog linear (ΔX, ΔY and ΔZ-axis) and angular deviations (ΔANGLE) were assessed. Euclidean distance (ΔEUC) was calculated from the linear deviation, reporting a mean of 82 µm (SD 61) ranging from 8 to 347 µm. ΔANGLE error mean was 0.33° (SD 0.20), ranging from 0.02 to 0.92°. From a multivariate analysis, when ΔEUC was considered as a response variable, a significant influence of ISB wear by scan number in interaction to position for implant 3.6 was identified (p < 0.0001); when ΔANGLE was considered as a response variable, a significant effect of position 3.6 was recorded ((p < 0.0001). The obtained results showed that the ISB wear negatively influenced the accuracy of IOS, suggesting that ISB base wear could be detrimental for the seating of ISBs on angulated implants.
Tipologia CRIS:
1.1 Articolo in rivista
Keywords:
accuracy; complete arch; digital impression; intraoral optical scanning; scanbody; wear
Elenco autori:
Arcuri, L.; Lio, F.; Campana, V.; Mazzetti, V.; Federici, F. R.; Nardi, A.; Galli, M.